rigas ONE
rigas-ONE developed by RIGAS Co., Ltd is a compact high pressure standard gas.
It is made to be easy to use anywhere at anytime. RIGAS provides customers with the various components
and concentrations needed for the calibration and so on.
PRODUCTS
Pure gas
Non-reactive gas
Automobile Exhaust Gas
Reactive gas(single component)
Reactive gas(multi components)
COMPACT
High pressure
REFILLABLE
PAMS, TO-14A
PAMS 57 Components mix (Ozone Precursor) - Provides R2
VOCs measured by the Photochemical Assessment and Measurement Station (PAMS) as precursors contributing to ozone generation
TO-14A 43 Components mix (Toxic Organics) - Provides R2
VOCs according to the US Environmental Protection Agency (EPA) standards as hazardous air pollutants in
the atmosphere
Special features
- Highly convenient and portable small 1L container (R2)
- Highly reliable values proven through short-term/long-term stability assessment
- KS I ISO 6142: Manufactured in accordance with 2015 standards (Gas analysis-Production of calibration gas mixture-Gravimetric method)
- Cylinders with special internal treatment
- Validity period of 12 months
- Precise measurement of components with low-vapor pressure and additional cylinder heating/regulator heating devices
Product information
Product |
Concentration |
Balance |
Specification |
PAMS |
5 nmol/mol |
N2 |
Blend tolerance: ± 30 % ; Analytical accuracy: ± 20 % |
1 μmol/mol |
N2 |
Blend tolerance: ± 10 % ; Analytical accuracy: ± 5 % |
TO-14A |
10 nmol/mol |
N2 |
Blend tolerance: ± 20 % ; Analytical accuracy: ± 10 % |
1 μmol/mol |
N2 |
Blend tolerance: ± 10 % ; Analytical accuracy: ± 5 % |
Analytical Tips
-
Some of the components with the low-vapor pressure in PAMS and TO-14A are gradually condensed in
cylinders depending on time elapsed, storage temperature, and usage environment after manufacturing.
As a result of it, lower concentration may be detected. Cylinder heating systems provided by RIGAS will
help you to use those components at the correct concentration.
-
Components with low-vapor pressure or strong adsorption may be adsorbed during the analysis process.
It will cause difficulty to detect accurate concentration. With regulator heating systems provided by RIGAS,
the concentration stabilization time of the components can be shorten and ensure accurate figures.
* Comparison of the peak size of n-Dodecane 1 μmol/mol in PAMS by time (3 months after manufacturing)
[Cylinder Heating Device]
Tips |
Cylinder Heating Device |
Regulator Heating Device |
Result |
|
Y |
Y |
Normal level |
|
Y |
N |
90 % of normal level |
|
N |
Y |
50 % less than normal |
|
N |
N |
50 % less than normal |
RIGAS supplies cylinder heating and regulator heating devices for precise component concentration analysis.
[Regulator Heating Device]